Irradiation at TIT 2004/04/05-06 -------------------------------- Rates @0cm 118kRad/h @10cm 61.3kRad/h @20cm 16.6kRad/h @24cm 11.5kRad/h @30cm 7.36kRad/h @70cm 1.36kRad/h Pixel irradiation ----------------- 2M sample: 17h@118kRad/h (with aminogray 9602241/209) 200k sample: 17h@11.5kRad/h from 2004/04/05 12:15 to 2004/04/06 05:33 PIN irradiation --------------- Amp. feedback setup a(board)b(pin)c(switch) any any 0 = 100MOhm 111 9.07M 112 0.98M 113 99.8k 121 9.04M 122 0.98M 123 99.7k 211 9.3M 212 0.98M 213 99.7k 221 9.01M 222 0.98M 223 99.9k Darwin Amp. Rate Total Volt. Material ch11 pin111 61.3 1042 0.29 3mm Al ch12 pin121 61.3 1042 0.29 / ch13 pin210 61.3 1042 3.4 / ch14 pin221 61.3 1042 0.33 / 2004/04/05 12:15 source up 2004/04/06 05:33 source down Material test 2004/04/05 10:21-10:26 10:30-10:32 10:34-10:59 11:07-11:57 12:01-12:18 ch11 110/3mm Al 3mm Al+2mm Pb - 4mm Pb - 2.93 2.58 3.05 2.42 3.02 ch12 120/1mm Pb 3mm Pb - 4mm Pb - 3.04 2.64 3.15 2.52 3.13 ch13 212/- 210/- 36.7e-3 0.232 ch14 223/- 220/- 8.09e-3 0.249 At 10:32 aminogray 9602241/210 removed from PIN11X board at 10cm Amplifier board exposed to 61.3kRad/h all the time, for total 5+2+25+30+17=1h19m i.e. about 100kRad ch11-12 always at 61.3kRad/h ch13-14 at 61.3kRad/h until 10:59, then at 1.36kRad/h second row are corresponding voltages in [V] PIN gain changes ---------------- When amplifier chip is >1m away from the source (dose rates below 1kRad/h) the gain change of the PIN system is small, 1-10% per MRad, depending on the PIN current. Gain always decreases. When Amp. board (with TL082CP operational amplifier) itself is exposed to high radiation: - noise and spikes largely increase - gain change becomes either positive or negative - it's magnitude can be up to 100% in one hour