The TOF and TSC scintillators (BC408, Bicron) were wrapped with one
layer of 45 m thick polyvinyl film (Tedlar) for light tightness
and surface protection. This thin wrapping minimizes the dead space
between adjacent TOF counters. Careful measurements of the physical
dimensions of all counters were made to ensure the dimensions of each
counter consistent with the TOF assembly tolerance.
The attenuation length and light yield of all of the TOF scintillators
were measured using a cosmic-ray test stand. A lucite Cerenkov
counter was used to suppress low momentum cosmic rays.
Fig. (a) shows the result of attenuation
length measurements for some of TOF scintillators. The attenuation
length was evaluated by using a ratio of signal charges and a time
difference between two PMTs of a TOF counter. The measured light
propagation velocity was 14.4 cm/ns. The average attenuation length
was 3.9 m, which satisfies the specification for the minimum
attenuation length of 2.5 m.
Fig. (b) shows the result of measurements of
light yields seen by each PMT for cosmic rays at the center of a
counter.
Figure:
(a) Attenuation lengths and (b) light yields of 38 TOF scintillators.
|