Next: ACC Readout Electronics
Up: Detector Design
Previous: Quality of the aerogels
  Contents
Since ACC is placed in a high magnetic field of 1.5 T in the Belle
detector, we decided to use fine-mesh photomultiplier tubes (FM-PMTs)
for the detection of Cerenkov lights, taking advantage of its
large effective area and high gain [44]. Other
candidate photo-sensors such as microchannel-plate photomultiplier
tubes (MCP-PMT) and hybrid photodiodes (HPD) were still at R&D stages
or extremely expensive when this decision was made. The FM-PMTs were
produced by Hamamatsu Photonics [49].
A sectional view of an FM-PMT is shown in
Fig. . Each FM-PMT has a borosilicate glass
window, a bialkali photo-cathode, 19 fine-mesh dynodes, and an
anode. Three types of FM-PMTs of 2, 2.5, and 3 inches in diameter are
used in ACC. The effective diameters () of these FM-PMTs are
39, 51, and 64 mm. The cathode-to-anode distance () is about 20
mm. The average quantum efficiency of the photo-cathode is 25 % at
400 nm wavelength. The optical opening of the mesh is about 50 %.
Figure:
Sectional view of FM-PMT.
|
The FM-PMTs with 19 dynode stages of fine mesh have high gain () with moderate HV values ( 2500 V). The gain of FM-PMT
decreases as a function of field strength as shown in
Fig. . The gain reduction is for
FM-PMTs placed parallel to the direction of magnetic field and
slightly recovers when they are tilted. The FM-PMTs used for ACC were
produced with dynodes with a finer mesh spacing than conventional
products at that time, to give approximately 10 times higher gain as
shown in in Fig. .
Figure:
Relative gains of conventional and improved FM-PMTs in
magnetic fields. The tubes were placed parallel (0) or tilted by
30 with respect to the field direction.
|
Effects of a magnetic field on the pulse height resolution have been
evaluated by tracing the change of a quantity , which is
defined as
, by using the mean () and
sigma () of the recorded ADC spectrum. The quantity
represents the effective photostatistics for the spectrum, which is a
convolution of a single-photoelectron (pe) response of the device and
the Poisson statistics with , the average number of
photoelectrons emitted from the photo-cathode. For FM-PMTs, the ratio
is about 2 at = 0. The relatively large excess
noise factor is due to the fact that a single-photoelectron spectrum
does not have any characteristic peak [50].
Reduction of in a magnetic field has been measured for 2
in., 2.5 in. and 3 in. FM-PMTs, with at = 0
() of about 20.
Table shows the ratio
at = 1.5 T for
two ranges of applied HVs. As the field increases, the resolution
deteriorates and the quantity decreases. The decrease in
is larger when the field is at a large angle such as 30
than the case of , and the decrease in at a
large angle is more significant for smaller FM-PMTs. However, it is
noticed that, in a magnetic field, the resolution improves, namely the
quantity increases, by applying higher voltage, while no
significant change is found in absence of the field.
Table:
Measured ratio
at B = 1.5 T. The number
of tested samples for each condition is shown in parentheses.
PMT |
2000 - 2200 V |
2600 - 2800 V |
diam. |
|
|
|
|
|
= 0 |
= 30 |
= 0 |
= 30 |
2 |
0.74 (7) |
0.64 (10) |
0.75 (10) |
0.73 (10) |
2.5 |
0.80 (5) |
0.73 (5) |
0.85 (8) |
0.85 (8) |
3 |
0.79 (2) |
0.77 (4) |
0.87 (6) |
0.99 (7) |
Next: ACC Readout Electronics
Up: Detector Design
Previous: Quality of the aerogels
  Contents
Samo Stanic
2001-06-02